Arkray Glucocard X-Meter Manuel d'instructions Page 12

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 72
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 11
12
These capillary action test strips require only

application tip for easy blood application.
The GLUCOCARD Vital Blood Glucose
Monitoring System measures the amount
of glucose (sugar) in fresh capillary whole
blood. When a drop of blood is added to the
sample application tip, the sample is drawn
into the test strip reaction site. The glucose
in the blood reacts with the enzyme on the
test strip. This reaction produces a current,
proportional to the amount of glucose in the
blood. The meter detects the current and
converts it into a blood glucose reading.
TEST STRIP
Reaction
Site
Test Strip
Contact Bars
Sample Application Tip
(apply blood to end of test strip
and capillary action will pull blood
to reaction site quickly and easily)
GLUCOCARD Vital TEST STRIPS
Vue de la page 11
1 2 ... 7 8 9 10 11 12 13 14 15 16 17 ... 71 72

Commentaires sur ces manuels

Pas de commentaire