User Instruction Manual
Read this User Instruction Manual
carefully before you start testing.
Vue de la page 0
1 2 3 4 5 6 ... 71 72

Résumé du contenu

Page 1 - User Instruction Manual

User Instruction Manual Read this User Instruction Manual carefully before you start testing.

Page 2 - CAUTION:

10FrontBackDisplay ScreenBack ButtonForward ButtonBattery Compartment CoverTest Strip PortUNDERSTANDING THE GLUCOCARD Vital BLOOD GLUCOSE

Page 3 - TABLE OF CONTENTS

11Explanation of the Full Display ScreenTimeDateAverage Test ReadingControl Solution TestLowBatteryUnit of MeasurementTemperature WarningTest StripApp

Page 4

12These capillary action test strips require only application tip for easy blood application. The GLUCOC

Page 5

13Important Test Strip Information  

Page 6 - INTRODUCTION

14Important Test Strip Information strips after 90 days (3 months) 

Page 7 - Intended Use

15Note: Changing the battery should only take a few seconds. If the battery is changed within 30 seconds, the meter memory is not affected. You will n

Page 8 - Important Information

16When the battery power is getting low, your meter shows . You have enough power for about 100 tests before you need to replace the battery.If the

Page 9 - THE GLUCOCARD Vital SYSTEM

17How to Change the BatteryStep 1Turn the meter off. The battery cover is on the back of the meter.Step 2Pull the battery cover up with your finger or

Page 10

18Meter Set UpBefore using the GLUCOCARD Vital Meter for the first time, you must set the 

Page 11

19Step 3: Set the time format the current time

Page 12 - GLUCOCARD Vital TEST STRIPS

2CAUTION: Before using any product to test your blood sugar (blood glucose), read all instructions and practice the test. Do all quality control check

Page 13

20Step 5: Set the hourPress and release the Back or Forward button to change the hour. Hold the Back or Forward button down to scroll through quickly.

Page 14

21Note: 

Page 15 - Changing the Battery

22Use the Assure Dose Control Solution to check that the meter and test strips are working correctly. It is important that you carry out this simple c

Page 16 - WARNING:

23It is recommended to use Control Solution:  

Page 17 - How to Change the Battery

24Performing a Control Solution TestBefore performing a control solution test, always check the expiration date ( ) of the control solution. Do

Page 18 - Setting the Time and Date

25Step 1Firmly insert a GLUCOCARD Vital Test Strip into the meter. Insert the contact bars into the test strip port, with the reaction site facing up.

Page 19 - 

26Step 2You must press the Back or Forward button BEFORE performing control solution test. Hold the Back or Forward button until appears in the up

Page 20 - Month flashing

27Note: Dye added to control solution may stain surfaces or fabric. Use with care.Step 5Compare the result with the range printed on the test strip bo

Page 21 - Year flashing

28Comparing Control Solution Results If the result is outside the range printed on the test strip bottle, repeat the 

Page 22 - ® DOSE CONTROL SOLUTION

29Repeat the test, carefully following instructions, using control solution mode with new control solution or a new test strip, as appropriate. Compar

Page 23

3Introduction ...6 Intended Use ...7 Important Information ...

Page 24

30Step 2 Insert a new, unused lancet into the Holder. Firmly press lancet until it stops and fits securely within the Holder. HolderMulti-Lancet Devic

Page 25

31Step 3 Twist off the protective cap.Step 4 Attach the removable cap.NOTE: Do not use the lancet if the cap is missing or damaged. A lancet should

Page 26

32Step 5 or thin skin or first time use, select “1 or 2.” If the obtai

Page 27 - 200-300

33Obtaining a Drop of BloodStep 1Wash hands with soap and warm water. Dry hands thoroughly. If you use alcohol wipes to cleanse fingers, make sure th

Page 28

34Step 6 to form. Keeping the hand warm, lowering the hand to waist level a

Page 29

35<<Note: For Instructions on Alternate Step 1Insert a GLUCOCARD Vital Test Strip into the meter. Insert the

Page 30 - Inserting a Lancet

36The test strip and flashing blood drop tells you the meter is ready for you to apply a drop of blood to the sample application tip at the end of the

Page 31

37not start the meter or can start the meter count down and provide an “E13” error

Page 32 - 1 2 3 4 5 6 7

38Step 3information such as insulin dosage, diet, and exercise. The result is

Page 33 - Obtaining a Drop of Blood

39Important Information About AST Sites other than your fingertip may have fewer nerve endings so obtaining a blood sample from these sites may be less

Page 34 - Disposing the Lancet

4Blood Glucose Testing ...30 The Lancing Device ...30 Inserting a Lancet ...

Page 35 - Test Strip Port

40Use Fingertip When Testing:   

Page 36 - INCORRECT

41Ask your diabetes healthcare professional about recommended testing procedures before operating machinery or driving a car, as fingertip testing is u

Page 37

42Step 2a) Insert a test strip into the meter. Insert contact bars into test strip port with reaction site facing up. When the strip is inserted, the

Page 38

43b) Wait until strip with flashing blood drop appears. (You now have 2 minutes to apply blood drop before meter shuts off).Step 3a) Vigorously rub the

Page 39 - ALTERNATE SITE TESTING (AST)

44Step 4Pick up the meter. Touch the end of the test strip to the blood sample until the reaction site is full. Immediately remove the meter and test

Page 40

45Step 5The meter will count down and display result in 7 seconds.CONSIDER YOUR RESULT.REPEAT ALTERNATE SITE TEST:

Page 41

46If the repeated alternate site result is still not consistent with how you feel, confirm your blood glucose level with fingertip testing.Step 6

Page 42 - Application Tip

47NOTE: The GLUCOCARD Vital Meter is designed to give results at 

Page 43

48Before You ComparePerform a control solution test to make sure your meter and test strips are working correctly. Wait at least 2 hours after a meal

Page 44 - IMPORTANT

49You may still experience a difference from the laboratory result because blood glucose levels can change over short periods. This can be for a varie

Page 45 - IF NOT, REPEAT TEST

5 Diabetes Findings ...51Meter Memory ...54 Marking Results in the Memory .

Page 46

50Expected Blood Glucose ReadingsExpected blood glucose levels for people without diabetes:1, 2 High Blood Glucose Readings

Page 47 - UNDERSTANDING TEST RESULTS

51Low Blood Glucose Readingstest with a new test strip. If this message s

Page 48

52What This Means for YouFrequent blood glucose testing is the best means you have for keeping track of 

Page 49

53References1 “Denitionanddiagnosisofdiabetesmellitusandintermediatehyperglycaemia”WorldHealthOrganization,20062 DiabetesCare,vol.3

Page 50 - High Blood Glucose Readings

54Marking Results in the MemoryYou can mark results in the meter memory in two ways.Marking Control Solution TestsYou must mark results with the contr

Page 51 - Diabetes Findings

55Marking Other ResultsYou can also mark results with an when you do NOT want to include the result 

Page 52

56Step 2Press the Back or Forward button to display the average of tests over the last 14 days. This can include 1 to 250 results, depending on how fr

Page 53

57< < mg/dL If a test has not been performed in the last 14 days the meter means there

Page 54 - METER MEMORY

58Results marked with when the test was performed are not included in the average. The marked results can be seen when you scroll through the res

Page 55 - Recalling Results

59A list of all display screen messages is included below.If you have any questions after reading these messages, or if the meter displays a message n

Page 56 - 

6INTRODUCTIONThank you for choosing the GLUCOCARD Vital Blood Glucose Monitoring System. Frequent blood glucose testing is important. It tells you how

Page 57

60The battery has been replaced.Check that the time and date are set correctly.The battery power is getting low. There is enough power for approximate

Page 58

61The battery power is low. There is only enough power for about 20 more tests. Replace with a new CR2032 battery as soon as possible.The battery powe

Page 59 - DISPLAY SCREEN MESSAGES

62meter and st

Page 60

63Corrupted memory.Call Customer Service at 800.566.8558.Retest with new test strip. Apply at least 0.5µL of

Page 61

64Cleaning the MeterThe GLUCOCARD Vital Blood Glucose Meter is a precise instrument. Please handle it with care. Clean the outside of the meter with a

Page 62

65  such as a micr

Page 63 - Corrupted memory

66TEST STRIPS: GLUCOCARD VITALRESULT RANGE:   CALIBRATION: SAMPLE SIZE: µLBLOOD SOURCE: CAPILLARY

Page 64 - Cleaning the Meter

67 substitute for pathology laboratory equipment and should not be used for

Page 65

68IMPORTANT: Loworhighbloodglucosereadingscanindicateapotentiallyseriousmedicalcondition.Ifyourbloodglucosereadingisunusuallylowo

Page 66 - SPECIFICATIONS

69THE GLUCOCARD Vital BLOOD GLUCOSE METERYou are a valued customer of ARKRAY. It is important to us that you are completely satisfied with your GLUCOCA

Page 67

7Intended UseThe GLUCOCARD Vital Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole bloo

Page 68

70THE LANCING DEVICEARKRAY warrants that the Lancing Device functions effectively for a period of 3 years from the date of purchase. A malfunctioning

Page 70 - THE LANCING DEVICE

©2011 ARKRAY, Inc.GLUCOCARD® is a registered trademark of ARKRAYARKRAY USA, Inc.Minneapolis, MN 55439 USATEL 800.566.8558FAX 952.64

Page 71

8Important Information   

Page 72 - ARKRAY USA, Inc

9The GLUCOCARD Vital System Includes:       